Prepare calibration equipment:

  • A MSA Altair 4XR gas detector (Normal operation)
  • Gas cylinder with correct type of gas inside
  • Valve regulator
  • Tubing (normally, supply with gas detector)
  • Calibration cap (normally, supply with gas detector)

Processing calibration:

  • Step 1: Assemby valve regulator and tubing to Gas cylinder
  • Step 2: Active calibration function on the MSA Altair 4XR by press and hold Up buttom within 3s
  • Step 3: Detector ask for confirm Zero calibration, press Power buttom to confirm
  • Step 4: After pass Zero calibration, detector ask for confirm Span calibration, press Power buttom to confirm
  • Step 5: Assembly calibration cap and open valve regulator then waiting auto calibration
  • Step 6: After Span calibration pass, close valve regulator, remove calibration cap then done the job.

If need more information, please call Hotline: 033 44 13586