Calibration MSA Altair 4XR
Prepare calibration equipment:
- A MSA Altair 4XR gas detector (Normal operation)
- Gas cylinder with correct type of gas inside
- Valve regulator
- Tubing (normally, supply with gas detector)
- Calibration cap (normally, supply with gas detector)
Processing calibration:
- Step 1: Assemby valve regulator and tubing to Gas cylinder
- Step 2: Active calibration function on the MSA Altair 4XR by press and hold Up buttom within 3s
- Step 3: Detector ask for confirm Zero calibration, press Power buttom to confirm
- Step 4: After pass Zero calibration, detector ask for confirm Span calibration, press Power buttom to confirm
- Step 5: Assembly calibration cap and open valve regulator then waiting auto calibration
- Step 6: After Span calibration pass, close valve regulator, remove calibration cap then done the job.