Zero calibration

When We need Zero Calibration?

  • Use a gas detector for detect over measuring range. Gas sensor may poisoned.
  • Test GasAlertMicroClip XL by non-gas standard ie: Lighter, cigarette, gas kitchen….

Where can be doing Zero Calibration GasAlertMicroClip XL

  • In fresh air; No flammable gas, no toxic, Oxygen concentration around 20.9%VOL

Performance:

  • Turn ON unit, let it go to normal operation
  • Press and hold Power button until unit Turn OFF, keep holding Power button until unit countdown to Calibration function.
  • Gas detector will be Auto Zero within 10s
  • Then detector required “Apply gas”, Press button for ignore it. (Because do not have gas cylinder, IF HAVE GAS CYLINDER, FULL CALIBRATION IS RECOMMEND)

Then, We done Zero Portalbe Multi-gas detector GasAlertMicroClip XL

See Full Calibration at: https://anhvucorp.com/category/calibration-instruction-videos/?lang=en