| Detector |
LEL: Catalytic bead, built in correction factor library |
| Oxygen: Liquid electrolyte |
| Toxic: Electrochemical CO, H2S, NH3, Cl2, NO2, SO2, HCN, NH3, PH3 |
| Battery |
Rechargeable Lithium-ion |
| Operating Periods |
Diffusion, Non-Wireless: Up to 14 hours |
| Pumped, Non-Wireless: Up to 11 hours |
| Diffusion, Wireless: Up to 10 hours |
| Pumped, Wireless: Up to 8 hours |
| Note: All battery specifications at 68°F, lower temperatures and alarm conditions will affect runtime |
| Graphic Display |
Monochrome graphic display (128×80) |
| Display Size: 1.57″ x 1/06″ with backlighting |
| Automatic or on-demand screen rotation |
| Keypad |
Two button operation |
| Direct Readout |
Real time reading of gas concentrations |
| Battery status |
| Pump status, if equipped |
| Wireless on/off and wireless reception quality |
| STEL, TWA, peak and minimum values |
| Man Down and policy enforcement indicators |
| Alarm |
Multi tone 95 dB buzzer, vibration alarm and flashing red LEDs and on-screen indication of alarm conditions |
| Alarms: Latching, non latching or manual override |
| Additional diagnostic alarm and display message for low battery |
| Pump stall alarm, pump version only |
| Man Down Alarm with pre-alarm and real time remote wireless notification |
| IP Rating |
Pumped: IP-65 / Diffusion: IP-67 |
| Data Logging |
Continuous datalogging, 3 months for 4 sensors at 1 minute intervals / User configurable datalogging intervals, from1 to 3,600 seconds |
| IP Rating |
Pumped: IP-65 / Diffusion: IP-67 |
| Calibration |
Two point calibration for zero and span (manual or automatic with AutoRae 2) |
| Sampling Pump |
Built in pump or diffusion / Can sample through tubing up to 98 feet |
| Temperature |
-4 to 122°F |
| Humidity |
0 to 95% rH, non condensing |
| Size |
Diffusion: 5.5″ L x 3.2″ W x 1.5″ H / Pumped: 5.7″ L x 3.2″ W x 1.7″ H |
| Weight |
Diffusion: 12.9 ounces / Pumped: 14.5 ounces |
| Approvals |
US & Canada: Class I, Division 1, Groups A, B, C, D |
Đánh giá